This paper discusses the characteristics of newly developed glucose biosensors with glucose oxidase deposited under high electric field electrophoresis. Enzyme deposition has been verified using electrochemical impedance spectroscopy. The response time of the fabricated sensor has been found to be within 10s to 20s. The developed sensor shows a sensitivity of 0.13μA/cm2/mM and a linear range of 0-12mM.