CQUniversity
Browse

Characteristics of glucose biosensors with glucose oxidase deposited under high electric field

conference contribution
posted on 2017-12-06, 00:00 authored by Daluwathu PreethichandraDaluwathu Preethichandra, Mala Ekanayake, K Kaneto
This paper discusses the characteristics of newly developed glucose biosensors with glucose oxidase deposited under high electric field electrophoresis. Enzyme deposition has been verified using electrochemical impedance spectroscopy. The response time of the fabricated sensor has been found to be within 10s to 20s. The developed sensor shows a sensitivity of 0.13μA/cm2/mM and a linear range of 0-12mM.

Funding

Category 3 - Industry and Other Research Income

History

Start Page

1953

End Page

1956

Number of Pages

4

Start Date

2012-01-01

Finish Date

2012-01-01

ISBN-13

9781457717710

Location

Graz, Austria

Publisher

IEEE Explore

Place of Publication

NJ, USA

Peer Reviewed

  • Yes

Open Access

  • No

External Author Affiliations

Faculty of Sciences, Engineering and Health; Institute for Resource Industries and Sustainability (IRIS); Kyūshū Kōgyō Daigaku; School of Engineering and Built Environment;

Era Eligible

  • Yes

Name of Conference

IEEE International Instrumentation and Measurement Technology Conference

Usage metrics

    CQUniversity

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC